model VCM-1500N2Space saving and inexpensive model
model VCM-780N2
■Sputum Collection Booth VCM-1500N2Although sputum examination is extremely important for a diagnosis of tuberculosis and evaluation of the outcome, airborne droplets generated at sputum collection should be prevented.
model VCM-1500N2
■Features
- Keeping the booth negative pressure prevents the aerosol generated at sputum collection from leaking out to the environment.
- Ventilation of 250 times/hour keeps the booth clean.
- Laminar air flow from the ceiling of the booth does not cause air entrainment, back-flow, or retention inside the booth.
- Air is supplied through a conventional HEPA filter, and exhausted through a HEPA filter treated with bactericidal enzymes.
- After use, inside the booth is disinfected with an UV lamp.
■Specifications
Model VCM-1500N2 Outer dimensions (mm) 1,500W × 900D × 2,175H Inner dimensions (mm) 1,150W × 794D × 1,823H Air flow system Down-flow system Cleanliness ISO Class 5 (Fed STD Class 100) Internal pressure Negative Filter air supply Dust collection efficiency of HEPA filter:PAO particle 0.3μm, 99.99% or more
Prefilter: nylon non-woven fabricexhaust Dust collection efficiency of HEPA filter:PAO particle 0.3μm, 99.99% or more
with bactericidal enzymes.
Prefilter: PP honeycomb net with bactericidal enzymes (option)Air flow rate 7.5 ㎥/min Ventilation frequency 250 times/h Fluorescent lamp 30W×1 UV lamp 15W×1 Power requirement Single phase 100V 50/60Hz 15A
■Sputum Collection Booth VCM-780N2
model VCM-780N2
■Sputum Collection Booth VCM-780N2
Tuberculosis (TB) is spread through the air when people who have an active TB infection cough, sneeze, or otherwise transmit their saliva through the air. To prevent the spread of TB, isolation of the patient or special ventilation is required. The sputum collection booth has been developed to prevent TB from spreading while collecting sputum, by keeping the booth under negative pressure, cleaning the exhaust air with HEPA filters, and radiating UV light.
- Small-footprint model that can easily be installed in the corner of a waiting room.
- Inexpensive model with simple functions and operations
- Ventilation frequency of 250 times/h
- Prompt removal of contaminated aerosol from the booth.
- Negative pressure indication light at the front panel allows you to easily confirm the internal pressure.
- Prevents the aerosol generated while sputum collection from spreading by keeping the pressure negative.
- A HEPA filter with bactericidal enzymes is equipped in the exhaustion side.
- Prevents the contaminated aerosol from spreading through the exhaustion side.
- The body is coated with melamine resin (antibacterial coating) suitable for preventing bacterial adherence.
- The front door has a see-through window (frosted glass) that allows you to know about the situation in the booth.
- An ultraviolet germicidal lamp is installed in the booth. It is equipped with an interlock unit to prevent turning on UV light while a patient is in the booth, by using a photo sensor. The UV light automatically turns off 2 hours after lighting.
Model VCM-780N2 Body Steel sheet, melamine resin paint (baked finish) Outer dimensions (mm) 780W × 1,235D × 2,045H Inner dimensions (mm) 700W × 860D × 1,800H Air flow system Down-flow system Internal pressure Negative Filter air supply Dust collection filter (strong non-woven cotton + synthetic fiber)
Trapping efficiency: 98% at 10 μm particle size
Prefilter: nylon non-woven fabricexhaust Dust collection efficiency of HEPA filter:PAO particle 0.3μm, 99.99% or more
with bactericidal enzymes.Air flow rate 5 ㎥/min or more Ventilation frequency 250 times/h or more Fluorescent lamp 20W×1 UV lamp 15W×1 Power requirement Single phase 100V 50/60Hz 15A
■Example of installation
In a hallway corner
(W1,500×D900×H2,175mm)In a waiting room for febrile patients
(W900×D1,730×H2,410mm)A wheelchair ramp can be installed.