本文へスキップ

Nippon Medical & Chemical Instruments Co., Ltd. is a manufacturer specialized for environmental control technology.

E-mail  nk.trade@nihonika.co.jp

3-9, Tamatsukuri Motomachi, Tennoji-ku, Osaka 543-0014, Japan

Plant Growth Chamber <Standard model> [LED lamp]  Products

Nippon Medical & Chemical Instruments Co., Ltd. is a manufacturer specialized for environmental control technology.

E-mail  nk.trade@nihonika.co.jp

3-9, Tamatsukuri Motomachi, Tennoji-ku, Osaka 543-0014, Japan

        

Only Temperature Control/Temperature & Humidity Control (LED)

Models L(P)H-241PFP/DT-S(P) and L(P)H-411PFP/DT-S(P) are standard plant growth chambers, which can control environment including short-day treatment or long-day treatment depending on research fields. There are two control modes available. One is a changeover mode, changing between constant operation and day/night switching operation. The other is a program control mode. Three-position controller is used for temperature and humidity control, which provides energy-saving effects. A wide variety of chambers are available in sizes ranging from 240 liters to 410 liters.

 


model LH-411PFDT-SP

 


■Features

  • An operation touch panel with time display is mounted on the upper part of the main body, which allows easy operation.
  • Model L(P)H-□PFP/DT-S allows both constant operation at any temperature/light/ (humidity) and day/night switching operation.
  • Model L(P)H-□PFP/DT-SP allows a program control.
  • Five-sided illumination. The light is continuously dimmable to 20% and 0%.
  • Model LHs don’t need a drain hose or a tank because of a forced evaporation system inside the chamber.
  • Three-position controller is used for temperature and humidity control (proportional control for heating), which provides low running costs.
  • The inner door has energy-saving functions, being used as one door or two up-and-down doors as per usual.
  • A hole of 40 mm diameter for measurement is included as standard.
  • Humidity control system can be selected from “only humidification” or “only dehumidification” or “both humidification and dehumidification.”



■Applications

  • Plant cell/tissue culture, germination, acclimation, or growth test
  • Environmental test
  • Growth of rice, arabidopsis, etc.
  • Insect experiment
  • Cell culture
  • Environmental resistance test for plants
  • Storage test at constant temperature and at constant humidity (only Model LPH)


■Options

  • Table for a water tank (Model LPH), network observing system, 3-in-1 LED illumination, metal halide lamp, carbon dioxide removal unit.

■Safety devices

  • Overheat prevention: thermal fuse, overheat thermoswitch (variable), alarm buzzer.
  • Overcool prevention: overcool thermoswitch (variable), alarm buzzer.
  • Earth leakage breaker.
  • Protection device for refrigerator (overload relay)



■Model designation







■Specifications

240 L series
Model  Pink  LH-241PFPT-S   LPH-241PFPT-S LH-241PFPT-SP  LPH-241PFPT-SP 
 Yellow LH-241PFDT-S   LPH-241PFDT-S LH-241PFDT-SP  LPH-241PFDT-SP 
 Outer dimensions (mm) 760W × 726D × 1,767H 
 Inner dimensions (mm) 512W × 485D × 980H 
Control system Changeover between constant operation and day/night switching operation  Step program, built-in clock control (24 hours)
24 steps/5 patterns, with pattern link functions  
Repeat ━  1~99 times or unlimited 
 Control methods Three-position control method: Refrigerator and humidifier, ON/OFF control; heater, proportional control 
Temperature  +5~50℃±1℃ (15~50℃ when all lights on)
Humidity  ━  50~80%RH±10%RH
(15~45℃)
 ━  50~80%RH±10%RH
(15~45℃)
Illuminance 0 to 8,500 lx for pale pink LEDs and 0 to 15,000 lx for pale yellow LEDs
They can be continuously dimmed to 20% and 0%.
 Photon flux density 0 to 175 μmolm-2s-1 for pale pink LEDs and 0 to 200 μmolm-2s-1 for pale yellow LEDs 
Light source LED lamps (Plant Flec) for plant growth
Number of LEDs  20W LED x 4 lamps x 2 sides (on the ceiling and door), 40W LED x 4 lamps x 3 sides (on the back, right, and left sides) 
Shelf  5 shelves (adjustable)  
Refrigerator 200W 
Heater  400W 500W 400W 500W 
Humidifier ━   80W   ━  80W 
 Operation current (maximum)  8.0A   12.5A   8.0A   12.5A 
 Power requirement Single phase 100V 50/60Hz 15A E 
  Weight (kg) About 220 About 235  About 225  About 235 

400 L series
Model   Pink  LH-411PFPT-S LPH-411PFPT-S   LH-411PFPT-SP LPH-411PFPT-SP 
 Yellow  LH-411PFDT-S LPH-411PFDT-S   LH-411PFDT-SP LPH-411PFDT-SP 
 Outer dimensions (mm) 880W × 806D × 1,875H
 Inner dimensions (mm) 670W × 565D × 1,100H 
Control systemd Changeover between constant operation and day/night switching operation Step program, built-in clock control (24 hours)
24 steps/5 patterns, with pattern link functions 
Repeat ━  1~99 times or unlimited
 Control methods Three-position control method: Refrigerator and humidifier, ON/OFF control; heater, proportional control 
Temperature  +5~50℃±1℃ (15~50℃ when all lights on)
Humidity  ━  50~90%RH±10%RH
(15~45℃)
 ━ 50~90%RH±10%RH 
(15~45℃)
Illuminance 0 to 8,500 lx for pale pink LEDs and 0 to 15,000 lx for pale yellow LEDs
They can be continuously dimmed to 20% and 0%.
 Photon flux density 0 to 180 μmolm-2s-1 for pale pink LEDs and 0 to 200 μmolm-2s-1 for pale yellow LEDs 
Light source LED lamps (Plant Flec) for plant growth
Number of LEDs  20W LED x 6 lamps x 1 side (on the ceiling), 40W LED x 4 lamps x 4 sides (on the back, right, and left sides) 
Shelf  6 shelves (adjustable)  
Refrigerator 300W 
Heater 500W 
Humidifier  ━  80W  ━   80W 
 Operation current (maximum)  10.0A   13.5A   10.0A   13.5A 
 Power requirement Single phase 100V 50/60Hz 15A E
  Weight (kg)  About 280 About 290  About 280  About 290 

# Temperature sensor: platinum resistance temperature sensor, humidity sensor: capacitance measurement sensor, humidifier: ultrasonic humidifier.
# Power requirement does not include the power-outlet installed in the body.

# These chambers are designed to operate in ambient temperatures up to 30ºC. The specified performance may not be achieved depending on actual usage conditions.
# If lighting, temperature setting should be 15ºC or higher.
# The specified performance may not be achieved depending on the combination of temperature and humidity.
# Depending on the sample, drainage installation may be required.
# Temperature can be controlled between +5 and +50ºC.
# Power outlets inside the chamber have a maximum current capacity of 1A.
# When controlling humidity, the temperature should be set between 15 and 45ºC in Model 241/411.