■Natural light model (Temperature/Temperature & Humidity Control)
Model LH-240PFP/D-N is a plant growth chamber which can utilize natural light. This model uses LED lighting for plant growth, and is most suitable for the experiments using both artificial and natural light. Long-day treatment is performed by the illumination lamps equipped in four sides. Three-position controller is used for temperature and humidity control, which provides energy-saving effects. A wide variety of chambers are available in sizes ranging from 120 liters to 410 liters.
model LH-240PFD-N
Operation Touch Panel
■Features
- This model is equipped with energy saving LED lamps with suitable spectral distribution for plant growth.
- An operation touch panel with time display is mounted on the upper part of the main body, which allows easy operation.(N model)。
- Easy to observe the inside, with double insulated glass and 5-sided structure. Suitable for experiments of plants or insects.
- Model N can changeover between constant operation (temperature/illuminance [LH-□N], temperature/humidity/illuminance [LPH-□N]) and day/night switching operation.
- Three-position controller is used for temperature and humidity control, which provides energy-saving effects.
- Air circulation system allows good temperature distribution.
- The inner door can be used as one door or two up-and-down doors as per usual, which allows a wide opening.
- A hole of 40 mm diameter for measurement is included as standard (option for Model RD).
- Humidity control system can be selected from “only humidification” or “only dehumidification” or “both humidification and dehumidification” (Model LPH).
■Lamp color
- You can select the lamp color from a pale pink and a pale yellow.
- PFP: Pale Pink
- PFD: Pale Yellow
■Applications
- Plant cell/tissue culture, germination, acclimation, or growth test
- Environmental resistance test for plants
- Insect experiment
- Storage test at constant temperature and at constant humidity (only Model LP)
- Environmental test
■Options
- Inside power-outlet
- Humidifier
- Table for a water tank (Model LPH) (W300×D400×H650mm)
■Safety devices
- Overheat prevention: thermal fuse, overheat thermoswitch (variable), alarm buzzer.
- Overcool prevention: overcool thermoswitch (variable), alarm buzzer.
- Earth leakage breaker.
- Protection device for refrigerator (overload relay)
■Model designation
■Specifications
Model Pink LH-120PFP-RD LH-240PFP-N LPH-240PFP-N LH-410PFP-N LPH-410PFP-N Yellow LH-120PFD-RD LH-240PFD-N LPH-240PFD-N LH-410PFD-N LPH-410PFD-N Outer dimensions (mm) W635×D602×H1,519 W790×D663×H1,749 W790×D693×H1,749 W960×D744×H1,864 W960×D747×H1,864 Inner dimensions (mm) W380×D418×H738 W515×D485×H980 W515×D485×H980 W670×D565×H1,100 W670×D565×H1,100 Control system Constrant operation, day/night switching light Changeover between constant operation and day/night switching operation Temperature +5~50℃±1℃ Humidity Option ━ 50~80%RH(15~45℃) ━ 50~90%RH(15~45℃) Illuminance LED (Plant Flec) for plant growth Light source 20W×12 lamps、10W×4 lamps 40W×12 lamps、15W×4 lamps 40W×12 lamps、20W×4 lamps Refrigerator 150W 250W 300W Heater 300W 400W 500W 500W Humidifier Option ━ 400W ━ 80W Shelf 4 shelves 5 shelves 6 shelves Power consumption Single phase 100V 5.5A Single phase 100V 7A Single phase 100V 13A Single phase 100V 7.5A Single phase 100V 11A Power requirement Single phase 100V 15A Single phase 100V 15A Single phase 100V 15A Single phase 100V 15A Single phase 100V 15A Weight About 90kg About 125kg About 130kg About 150kg About 155kg
# Temperature sensor: platinum resistance temperature sensor, humidity sensor: capacitance measurement sensor, humidifier: steam humidifier (LPH-240PFP/D-N) or ultrasonic humidifier (LPH-410PFP/D-N)
# When LH-120RD has a humidifier, the humidity may become constant above the set value.
# These chambers are designed to operate in ambient temperatures up to 30ºC. The specified performance may not be achieved depending on actual usage conditions.
# Depending on operating conditions, samples in the chamber may become dry.
# When lighting, temperature setting should be 15ºC or higher.
# Power requirement does not include the power-outlet installed in the body.
# The specified performance may not be achieved depending on the combination of temperature and humidity.